Class II — Avanos Medical, Inc.
| Recalling firm | Avanos Medical, Inc. |
|---|---|
| Registered facility | View this company's 3 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | Avanos Medical has received reports stating that the central lumen of some 5 Fr Neonate/Pediatric Closed Suction Catheters may be fully blocked or partially blocked. This may cause the closed suction catheters (CSCs) to inadequately suction secretions and/or saline fluids form the patient's airway to prevent oxygen desaturation. |
| Product | Haylard Closed Suction System for Neonates/Pediatrics, 5 Fr, 1. REF 195, Y-Adapter, SAP Finished Product Code 109382303; 2. REF 201, with BALLARD Technology, SAP Finished Product Code 109382803; 3. REF 202, Elbow, SAP Finished Product Code 109382903; 4. REF 195-5 Y-Adapter, SAP Finished Product Code 109838302. |
| Quantity | 114,300 units (37,880 US, 76,420 OUS) |
| Distribution | Nationwide. Internationally distributed to: Argentina, Australia, Austria, Brazil, Canada, Chile, Columbia, Croatia, Czech Republic, Denmark, France, Germany, Greece, Honduras, India, Ireland, Israel, Italy, Korea, Republic of, Lebanon, Mexico, Nepal, Netherlands, New Zealand, Norway, Peru, Poland, Portugal, Russian Federation, Saudi Arabia, Singapore, Slovakia, Slovenia, Spain, Sweden, Switzerl… |
| Initiated | 2019-11-05 |
| Classified | 2020-01-08 |
| Terminated | 2023-08-16 |
| Recall number | Z-0788-2020 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Avanos Medical, Inc. (Z-0788-2020).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/84227/