Recall

Class II — Enztec Limited

Recalling firmENZTEC LIMITED
Registered facilityENZTEC LIMITED (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonThe femoral impactors have a higher than normal risk of the arm breaking. During impaction the tip of the arm may fracture and separate from the instrument. This may cause a minor delay in surgery. If not retrieved, it may cause an inflammatory response that may require revision surgery.
ProductREF 800-05-035, ENZTEC FEMORAL IMPACTOR - DJO, (01) 09421028110624 (10) 999999, To be used only with DJO EMPOWR 3D KNEE
Quantity312 Femoral Impactors
DistributionCA
Initiated2019-08-16
Classified2019-12-06
Terminated2023-03-02
Recall numberZ-0612-2020
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)