Class II — Cardinal Health 200, LLC
| Recalling firm | Cardinal Health 200, LLC |
|---|---|
| Registered facility | View this company's 35 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | The frequency of calibration for the Genius Tympanic Thermometer as stated in the operating manual may not ensure that thermometers always remain within the stated accuracy range ( 0.2C for Genius 2 and 0.3C for Genius 3 thermometers). The measurement readings drift upwards over time, which means that the thermometers could exceed the upper stated accuracy tolerance of +0.2C for Genius 2 or +0.3C for Genius 3. |
| Product | Genius 3 Tympanic Thermometer- electric thermometer Item Code: 303013 |
| Quantity | 7425 US; OUS 214278 |
| Distribution | Worldwide Distribution - US Nationwide and Foreign in the countries of: Australia Austria Belgium Canada Canary Islands Chile Costa Rica Croatia Cyprus Czech Republic Denmark Egypt Estonia Finland France French Guiana French Polynesia Gabon Germany Gibraltar Greece Guadeloupe Hong Kong Iran, Islamic Republic Of Ireland Israel Italy Kuwait Liechtenstein Luxembourg Mar… |
| Initiated | 2019-10-28 |
| Classified | 2020-02-04 |
| Terminated | 2022-05-17 |
| Recall number | Z-0991-2020 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Cardinal Health 200, LLC (Z-0991-2020).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/84202/
Alerts for Cardinal Health 200, LLC
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.