RegWardenFDA facility compliance records
Recall

Class II — LivaNova USA Inc

Recalling firmLivaNova USA, Inc.
Registered facilityLivaNova USA, Inc. (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonThis recall is being initiated due to reports that that the therapy programming tablet with software version 1.5 errantly performs a normal mode diagnostic test instead of the selected system diagnostic test on Model 102 and Model 102R devices, if the output current is greater than 0.5mA. This can result in false high impedance values during patient follow-up.
ProductVNS Therapy Program GMDN: 44077 Model # 3000 V1.5.2, RxOnly
Quantity643 units
DistributionIL, AK, AL, AR, AZ, CA, CA , CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD MI, MN, MO, MS, MT, NC, ND, NE, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WA , WI, WV, WY,
Initiated2018-11-16
Classified2019-12-10
Recall numberZ-0663-2020
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — LivaNova USA Inc (Z-0663-2020).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/84170/

Not open yet

Alerts for LivaNova USA, Inc.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.