Class II — LivaNova USA Inc
| Recalling firm | LivaNova USA, Inc. |
|---|---|
| Registered facility | LivaNova USA, Inc. (the firm's only registered facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | This recall is being initiated due to reports that that the therapy programming tablet with software version 1.5 errantly performs a normal mode diagnostic test instead of the selected system diagnostic test on Model 102 and Model 102R devices, if the output current is greater than 0.5mA. This can result in false high impedance values during patient follow-up. |
| Product | VNS Therapy Program GMDN: 44077 Model # 3000 V1.5.2, RxOnly |
| Quantity | 643 units |
| Distribution | IL, AK, AL, AR, AZ, CA, CA , CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD MI, MN, MO, MS, MT, NC, ND, NE, NE, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WA , WI, WV, WY, |
| Initiated | 2018-11-16 |
| Classified | 2019-12-10 |
| Recall number | Z-0663-2020 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — LivaNova USA Inc (Z-0663-2020).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/84170/