RegWardenFDA facility compliance records
Recall

Class II — Abiomed, Inc.

Recalling firmABIOMED, INC.
Registered facilityView this company's 2 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonThis one pump shipped with incorrect pump programming, the initial production working version of the program. The pump would have immediately run at P9 for all selected P-levels, rather than a 10 second ramp up period. The pump may have migrated out of position and possibly caused injury. The positioning and suction alarms were also disabled.
Productlmpella CP with Smart Assist, Model Number: 0048-0003. Left heart support blood pump, for Use During Cardiogenic Shock and High-Risk PCI
Quantity1
DistributionThe product was distributed to GA
Initiated2019-09-04
Classified2019-11-07
Terminated2020-01-21
Recall numberZ-0345-2020
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Abiomed, Inc. (Z-0345-2020).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/84025/