RegWardenFDA facility compliance records
Recall

Class II — GN Hearing A/S

Recalling firmGN Hearing A/S
Registered facilityView this company's 5 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonGN Hearing received reports from our manufacturing site in US regarding a software bug in our service & repair tool which affected the capture and restoration of user data in a limited number of repaired hearing aids. The software bug caused these repaired hearing aids to contain altered specifications for sound amplification, potentially putting users residual hearing at risk of experiencing unintended additional available gain to varying degrees i.a.18dB flat linear gain or potentially in worst case Full On Gain. However, no injuries have been reported.
ProductIn-the-Ear hearing aids, Model Numbers: T9ITC-DW-LP, FT6ITC-D-MP, FT6ITE-D-MP, FT8CIC-MP, FT8ITC-W-HP, FT8ITC-W-LP, FT8ITC-W-MP, FT8ITE-DW-HP, FT8MIH-W-HP, FT8MIH-W-MP, LT5CIC-MP, LT5CIC-UP, LT5IIC-LP, LT5ITC-DW-MP, LT5ITC-DW-UP, LT5ITE-DW-HP, LT5ITE-DW-MP, LT5ITE-DW-UP, LT5MIH-W-HP, LT5MIH-W-MP, LT5MIH-W-UP, LT7CIC-HP, LT7CIC-LP, LT7CIC-MP, LT7CIC-W-MP, LT7IIC-LP, LT7ITC-DW-LP, LT7ITC-DW-MP, LT7ITC-MP, LT7ITC-W-HP, LT7ITC-W-MP, LT7ITE-DW-HP, LT7ITE-DW-MP, LT7ITE-DW-UP, LT7ITE-W-UP, LT7MIH-W-HP, LT7MIH-W-LP, LT7MIH-W-MP, LT9CIC-HP, LT9CIC-MP, LT9CIC-W-MP, LT9IIC-LP, LT9ITC-DW-HP, LT9ITC-DW-L…
Quantity285 units
DistributionNationwide
Initiated2019-09-18
Classified2019-11-27
Terminated2023-05-31
Recall numberZ-0584-2020
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — GN Hearing A/S (Z-0584-2020).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/83906/

Not open yet

Alerts for GN Hearing A/S

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.