Class II — Siemens Medical Solutions USA, Inc

Recalling firmSIEMENS MEDICAL SOLUTIONS USA, INC.
Registered facilityView this company's 9 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonArtis zee/Q systems patient tables may potentially have cracks in the table mainframe, and cause mechanical detachment of the upper part of the patient bed and lead to hazardous situations for patients and medical personnel due to falling parts
ProductArtis zee and Artis Q/ Q.zen Interventional Fluoroscopic X-Ray System with patient tables as follows: Systems Material Number Artis zee biplane 10094141; Artis Q floor 10848280; Artis Q biplane 10848282; Artis Q.zen floor 10848353; Artis Q.zen biplane 10848355; Artis zee ceiling 10094137; Artis Q ceiling 10848281; Artis Q zeego 10848283 ArtisQ.zen ceiling 10848354 Product Usage: Artis is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional proc…
QuantityUS 17
DistributionUS Nationwide Distribution - AL, AR, KS, MI, MN, NC, ND, NY, OH, SC, TX, WI
Initiated2019-07-30
Classified2019-08-24
Terminated2020-04-29
Recall numberZ-2367-2019
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Siemens Medical Solutions USA, Inc (Z-2367-2019).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/83565/

Not open yet

Alerts for SIEMENS MEDICAL SOLUTIONS USA, INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.