RegWardenFDA facility compliance records
Recall

Class II — ICU Medical Inc

Recalling firmICU MEDICAL, INC.
Registered facilityView this company's 6 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonA software malfunction of the infuser when used with a MedNet Custom Drug Library (CDL). When a new CDL is downloaded and installed on the infuser the infuser can become inoperable and show a blank screen on the User Interface.
ProductPlum 360 Infusion System With MedNet/ Smart Card Plug And Play Module, Model # 30010. Infusion pump.
Quantity47,152 devices
DistributionUS nationwide distribution, and Canada
Initiated2018-08-01
Classified2019-10-22
Terminated2021-04-27
Recall numberZ-0171-2020
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — ICU Medical Inc (Z-0171-2020).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/83480/