RegWardenFDA facility compliance records
Recall

Class II — Luminex Corporation

Recalling firmLuminex Corporation
Registered facilityView this company's 3 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonThrough internal investigation, it was discovered that a portion of side cassette assemblies (SCAs) back side seal (BSS) did not meet design specification, resulting in partially sealed SCAs. The impact of a partial seal of the BSS could result in potential decrease of volume for each solution in the SCAs and decreased sensitivity or stability of the assay.
ProductARIES Group A Strep Assay , REF 50-10041, UDI # 00840487101469
QuantityN/A
DistributionUS: AK, AL, AZ,CA,CO,FL,GA,IL,KS,KY,LA,MA,MD,ME,MI,MN,MO,MS,MT,NC,NE,NJ,NM,NY,OH,ON,OR,PA,TN,TX,UT,VA,VT,WA,WI, OUS: Belgium, Canada, Thailand, Germany, Singapore, United Kingdom
Initiated2019-02-20
Classified2019-08-15
Recall numberZ-2286-2019
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Luminex Corporation (Z-2286-2019).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/83470/

Not open yet

Alerts for Luminex Corporation

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.