RegWardenFDA facility compliance records
Recall

Class II — CareFusion 303, Inc.

Recalling firmCareFusion 303, INC.
Registered facilityView this company's 3 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonCalibration; it has been determined that the capnographs may have been calibrated during manufacturing using a lower than specified concentration of CO2
ProductAlaris EtCO2, Model 8300, UDI: 10885403830013, CareFusion for models 11634567 and 8300ADXeN933
Quantity76 EtCO2 units
DistributionUS: KY, TX CA, IL, MN, NC, IA, SD, KS, DC OUS: Australia
Initiated2019-07-01
Classified2019-08-30
Terminated2020-09-09
Recall numberZ-2443-2019
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — CareFusion 303, Inc. (Z-2443-2019).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/83296/

Not open yet

Alerts for CareFusion 303, INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.