Class II — Elekta Inc
| Recalling firm | ELEKTA AB |
|---|---|
| Registered facility | View this company's 4 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Ongoing |
| Reason | When creating 3D plans using either MU or Dose weighting modes, if the user changes the number of fractions, rescales the plan and then changes the wedge angle, the Monitor Units are scaled incorrectly. |
| Product | Monaco RTP System, software versions 5.10 and 5.11, Radiation Treatment Planning System |
| Quantity | 1396 units |
| Distribution | worldwide |
| Initiated | 2019-07-03 |
| Classified | 2019-07-27 |
| Recall number | Z-2091-2019 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Elekta Inc (Z-2091-2019).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/83273/