RegWardenFDA facility compliance records
Recall

Class II — Teleflex Medical

Recalling firmTeleflex
Registered facilityView this company's 9 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonPotential for product not to be sterile.
ProductRusch EasyCath Kit
Quantity3,700 devices
DistributionNationwide distribution to CA, FL, IL, NC, NJ, OH, OK, TX.
Initiated2019-07-01
Classified2019-07-26
Terminated2021-11-04
Recall numberZ-2086-2019
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Teleflex Medical (Z-2086-2019).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/83250/

Not open yet

Alerts for Teleflex

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.