RegWardenFDA facility compliance records
Recall

Class II — Ad-Tech Medical Instrument Corporation

Recalling firmAD-TECH MEDICAL INSTRUMENT CORP.
Registered facilityAD-TECH MEDICAL INSTRUMENT CORP. (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonOn June 12, 2019 Ad-Tech was made aware that on three different occasions between March 1, 2019 and May 31, 2019 Ad-Tech personnel have distributed non-sterile anchor bolt parts to one independent representative and one international distributor, who may have subsequently sterilized the non-sterile anchor bolts for clinical use. Ad-Tech does not have regulatory clearance to distribute non-sterile anchor bolts and does not have a validated hospital sterilization method for these devices. Further, Ad-Tech labeling does not include instructions for sterilizing anchor bolts.
ProductLSB Style Anchor Bolt. Labeled with the following parts: 13mm anodized anchor bolt body with anchor bolt cap containing a blue gasket and silicone cap, 21mm anodized anchor bolt body with anchor bolt cap containing a blue gasket and silicone cap, 13mm non-anodized anchor bolt body with anchor bolt cap containing a green gasket and silicone cap, 26mm non-anodized anchor bolt body with anchor bolt cap containing a blue gasket and silicone cap - Product Usage: The Ad-Tech Anchor Bolts are optional accessories for use with Depth Electrodes. The Anchor Bolts may be applied when it is desired …
Quantity201 devices
DistributionWorldwide distribution - US Nationwide distribution in the state of Georgia and country of Australia.
Initiated2019-06-18
Classified2019-08-06
Terminated2021-02-16
Recall numberZ-2204-2019
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Ad-Tech Medical Instrument Corporation (Z-2204-2019).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/83246/