RegWardenFDA facility compliance records
Recall

Class II — Spacelabs Healthcare, Inc.

Recalling firmSPACELABS HEALTHCARE, INC.
Registered facilityView this company's 2 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonSeveral reports were received that patient records were printed with the correct patient demographics but containing another patient's waveforms.
ProductSpacelabs Healthcare Smart Disclosure System, Model 92810, a component of the Intesys Clinical Suite (ICS) G2.
Quantity295 customer sites have the affected software versions
DistributionDistribution was nationwide. There was also government and military distribution. Foreign distribution was made to Argentina, Bahrain, Bolivia, Brazil, Canada, Cayman Islands, Chile, China, Costa Rica, Czech Republic, France, Hong Kong, India, Indonesia, Italy, Kuwait, Malaysia, Mexico, Netherlands, Pakistan, Palestine, Panama, Philippines, Poland, Qatar, Romania, Saudi Arabia, Spain, Switzerl…
Initiated2019-01-28
Classified2019-07-22
Terminated2022-02-24
Recall numberZ-2035-2019
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Spacelabs Healthcare, Inc. (Z-2035-2019).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/83220/

Not open yet

Alerts for SPACELABS HEALTHCARE, INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.