RegWardenFDA facility compliance records
Recall

Class II — Randox Laboratories, Limited

Recalling firmRANDOX LABORATORIES, LTD.
Registered facilityView this company's 4 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonThere was a report of a malfunction with the sensor which resulted in an increase of temperature leading to melting of internal wires and scorch damage to the internal structure of the instrument.
ProductRX Daytona, Model Nos. RX4040 and RX4041
Quantity3
DistributionDomestic distribution to CA CT MA MD MO NC NJ NY OH PA UT WV and PR. Worldwide foreign distribution.
Initiated2019-06-24
Classified2019-07-25
Terminated2020-05-21
Recall numberZ-2083-2019
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Randox Laboratories, Limited (Z-2083-2019).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/83175/