Class I — Cook Inc.
| Recalling firm | COOK, INC. |
|---|---|
| Registered facility | View this company's 5 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class I |
| Status | Terminated |
| Reason | Multiple complaints were received for balloons bursting below the rated burst pressure. Potential adverse events that may occur if an affected product is used include a delay in the procedure, additional intervention, vessel injury, and balloon fragmentation in the patient. |
| Product | Advance Enforcer 35 Focal Force PTA Balloon Catheter 6mm x 4cm, 5FR/80cm, Catalog Number ASB5-35-80-6-4, REF Number G35252 Product Usage: The Advance Enforcer 35 Focal-Force PTA Balloon Catheter is intended for percutaneous transluminal angioplasty (PTA) of lesions in peripheral arteries, including iliac, renal, popliteal, infrapopliteal, femoral and iliofemoral, as well as obstructive lesions of native or synthetic arteriovenous dialysis fistulae. Not for use in the cerebral or coronary vasculature. |
| Quantity | 131 |
| Distribution | Worldwide Distribution - US Nationwide The affected product was distributed to the following states: FL, OH, WA. The products were distributed to the following foreign countries: Belgium, Denmark, Finland, Germany, Italy, Netherlands, Norway, Poland, South Africa, Spain, Sweden, Switzerland, United Arab Emirates, United Kingdom. |
| Initiated | 2019-05-24 |
| Classified | 2019-06-17 |
| Terminated | 2019-10-31 |
| Recall number | Z-1757-2019 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class I recall — Cook Inc. (Z-1757-2019).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/82888/
Alerts for COOK, INC.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.