RegWardenFDA facility compliance records
Recall

Class II — Polymer Technology Systems, Inc.

Recalling firmPOLYMER TECHNOLOGY SYSTEMS, INC.
Registered facilityPolymer Technology Systems (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonSome lots of test strips do not fit tightly into the optical block, and may fall out during analysis, which could lead to an incorrect result being displayed
ProductCHOL+HDL+GLU test strips, REF 2412
Quantity67643 total
DistributionDomestic distribution nationwide.
Initiated2019-01-16
Classified2019-05-23
Terminated2020-06-24
Recall numberZ-1373-2019
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Polymer Technology Systems, Inc. (Z-1373-2019).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/82331/

Not open yet

Alerts for POLYMER TECHNOLOGY SYSTEMS, INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.