RegWardenFDA facility compliance records
Recall

Class II — AAP Implantate Ag

Recalling firmAAP IMPLANTATE AG
Registered facilityAAP IMPLANTATE AG (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonThere is the potential for the Lambotte chisel to break at the welded piece parts of the chisel body and knocking plate.
Productaap Implante AG, Lambotte chisel, 20 mm, REF IU 3000-20 Product Usage: Surgical instruments are intended to aid in the insertion of bone plates and screws as well as for reduction and support during orthopedic surgeries. If not otherwise specified, instruments are intended for multiple use.
Quantity25 units
DistributionWorldwide Distribution - US Nationwide in the state of Texas and country of Germany
Initiated2019-01-11
Classified2019-03-17
Terminated2022-05-26
Recall numberZ-1019-2019
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — AAP Implantate Ag (Z-1019-2019).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/82159/

Not open yet

Alerts for AAP IMPLANTATE AG

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.