Recall

Class II — Beckman Coulter Inc.

Recalling firmBECKMAN COULTER, INC.
Registered facilityView this company's 22 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonWhen racks are front-loaded onto the sample presentation unit while samples from the automation line are being processed, a scheduling error within the software can cause mishandling of samples during the creation of an aliquot and lead to incorrect results being generated. Affected samples typically yield non-reproducible falsely decreased results by up to 100%, however the potential for falsely elevated results cannot be ruled out.
ProductUniCel DxC 660i, 680i, 860i, and 880i SYNCHRON Access Clinical Systems, that are re-configured so that the DxI system is separated from the DxC system and the DxI instrument is then connected to an automation line. System Name/ Catalog Number: DxC 660i/ A64871, DxC 680i /A64903, DxC 860i /A64935, DxC 880i /A59102
Quantity0
DistributionUS: NJ, OH, NY, GA, FL, TN, WI, MI, MS, CA, NC, TX, PA, DC, IN, KY, IL, WA, MN, LA, ND, ID, WV, AK, AL, MA. OUS: Argentina, Australia, Bahrain, Canada, China, Croatia, Czech Republic, Egypt, Germany, Greece, India, Italy, Malaysia, Morocco, Netherlands, Nigeria, Poland, Portugal, Qatar, Saudi Arabia, Singapore, South Africa, Spain, Switzerland, Thailand, Tunisia, United Kingdom, United States, …
Initiated2018-12-05
Classified2019-03-30
Recall numberZ-1078-2019
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Beckman Coulter Inc. (Z-1078-2019).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/82079/

Not open yet

Alerts for BECKMAN COULTER, INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.