RegWardenFDA facility compliance records
Recall

Class II — Merge Healthcare, Inc.

Recalling firmMerge Healthcare
Registered facilityView this company's 2 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonCertain configurations of Merge Eye Station hardware may result in electrical output that exceeds the allowable limits for medical devices. Leakage current was found to exceed allowable limits in earth leakage current, touch current in a single fault condition (of open ground) and patient leakage current in a single fault condition (of open ground).
ProductThe WinStation Retinal Imager, also known as Merge Eye Station. DCF-1024 & DCF-512 Digital Imaging System Product Usage: The WinStation Retinal Imager has the same intended use as other fundus cameras. It is specifically used to produce color or black and white images of the retina (fundus imaging) and anterior segement (slit lamp imaging), and fluorescein and ICG angiographic images in a non-invasive manner.
Quantity1,252 Units
DistributionDistribution nationwide to Alabama, Alaska, Arizona, California, Colorado, Connecticut, Delaware, District of Columbia, Florida, Georgia, Hawaii, Idaho, Illinois, Indiana, Iowa, Kansas, Kentucky, Louisiana, Maine, Maryland, , Massachusetts, Michigan, Minnesota, Mississippi, Missouri, Montana, Nebraska, Nevada, New Hampshire, New Jersey, New York, North Carolina, North Dakota, Ohio, Oklahoma, Orego…
Initiated2018-12-20
Classified2019-02-25
Terminated2023-02-28
Recall numberZ-0939-2019
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Merge Healthcare, Inc. (Z-0939-2019).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/81869/

Not open yet

Alerts for Merge Healthcare

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.