RegWardenFDA facility compliance records
Recall

Class II — Waldemar Link GmbH & Co. KG (Mfg Site)

Recalling firmWALDEMAR LINK GMBH & CO. KG
Registered facilityWaldemar Link GmbH & Co. KG (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonThe reaming guide and the guide rod can't be screwed into the intended depth. The reason for this is a reduced depth of the core hole in the stem. A reaming guide that is not screwed in completely may lead to a proximal preparation in the depth that is not sufficient and to a neck that can't be positioned correctly. The correct fit of the neck can't be verified by means of the guide rod.
ProductLINK STEM with Microporous Surface MP Reconstruction Prosthesis, CEMENTLESS (Size XS-0, D: 12mm, L: 160mm), Reference Number 172-916/12 Product Usage: The LINK® STEM with Microporous Surface MP® Reconstruction Prosthesis, CEMENTLESS System is indicated for patients with mobility-limiting diseases, fractures or defects which cannot be treated by conservative or osteosynthetic procedures. The LINK® STEM with Microporous Surface MP® Reconstruction Prosthesis, CEMENTLESS System is indicated for the following conditions: Revision arthroplasty due to juxta-articular bone defects Revision of …
Quantity11
DistributionUS Nationwide in the states of: FL, GA, IA, IN, KS, MI, MN, and TN.
Initiated2018-12-07
Classified2019-01-14
Terminated2019-12-17
Recall numberZ-0733-2019
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Waldemar Link GmbH & Co. KG (Mfg Site) (Z-0733-2019).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/81778/

Not open yet

Alerts for WALDEMAR LINK GMBH & CO. KG

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.