RegWardenFDA facility compliance records
Recall

Class II — Stihler Electronic Gmbh

Recalling firmSTIHLER ELECTRONIC GMBH
Registered facilitySTIHLER ELECTRONIC GMBH (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonPRISMAFLO II and PRISMAFLO II S Heating Profile only and not the Control Unit, damage may result in a hotspot on the Heating Profile, may cause low level skin burn
ProductHeating Profile PRISMAFLO II blood return warmer for Prismaflex Product Code: PF2-WP2618 Product Usage: The PRISMAFLO IIS warmer is used for warming return blood flow.
Quantity247 units
DistributionWorldwide Distribution - US Nationwide in the states of CA and Foreign of: Switzerland
Initiated2018-09-06
Classified2019-01-23
Terminated2020-05-13
Recall numberZ-0761-2019
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Stihler Electronic Gmbh (Z-0761-2019).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/81755/