Class II — Covidien LLC
| Recalling firm | Covidien llc |
|---|---|
| Registered facility | View this company's 22 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | The firm is issuing this voluntary removal following receipt of reports of parastomal mesh failure identified several years following parastomal hernia repair using the modified Sugarbaker repair technique. In these reports, product failure led to hernia recurrence requiring additional surgical treatment. Symptoms of hernia recurrence may include discomfort, localized pain-free or painful bulging, and possible changes in the overlying skin. Medtronic has received, worldwide, a total of ten reports of mesh failure following use of the product in the last five years. Patients who have received a Parietex" composite parastomal mesh for the treatment of a parastomal hernia should continue to receive ongoing monitoring by their healthcare providers for the recurrence of a parastomal hernia. |
| Product | Parietex Composite Parastomal Mesh (Product Number PCOPM15 / PCOPM20) Product Usage: The PCOPM15 and PCOPM20 reorder codes are circular in shape, made from a three-dimensional monofilament polyester fabric with a two-dimensional monofilament polyester central band. One side of the design is completely covered with the hydrophilic film. On the opposite side, only the two-dimensional central band is coated with the absorbable hydrophilic film. The film extends 5 mm over the external edge of the reinforcement. The Parietex composite parastomal mesh is indicated for the reinforcement of soft t… |
| Quantity | 7333 |
| Distribution | Worldwide distribution--US Nationwide |
| Initiated | 2018-10-05 |
| Classified | 2018-11-14 |
| Terminated | 2020-05-07 |
| Recall number | Z-0461-2019 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Covidien LLC (Z-0461-2019).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/81417/
Alerts for Covidien llc
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.