RegWardenFDA facility compliance records
Recall

Class II — Smith & Nephew, Inc.

Recalling firmSmith & Nephew, Inc.
Registered facilityView this company's 15 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonOne lot of the Legion AP femoral cutting block, size 6, was manufactured with an oversized spring.
ProductSmith+Nephew Orthopaedic Division LEGION A/P FEMORAL CUTTING BLOCK, SIZE 6 REF 71434411, QTY: (1) The LEGION AP Femoral Cutting Block is a reusable surgical instrument used to make bone cuts on the distal femur to prepare the bone to receive a LEGION femoral implant.
Quantity6 units
DistributionUS Distribution to the states of: KS and TN.
Initiated2018-09-18
Classified2018-10-22
Terminated2019-05-08
Recall numberZ-0200-2019
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Smith & Nephew, Inc. (Z-0200-2019).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/81100/