Class II — Vision RT Ltd
| Recalling firm | VISION RT LIMITED |
|---|---|
| Registered facility | View this company's 3 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | RTInterface error in the event of an I/O error between the IGC PCB and the Framegrabber PCB |
| Product | Integrated Gate Controller PCB (component of AlignRT Plus) Truebeam variant only. The AlignRT is a video-based three-dimensional (3D) surface imaging system which is used to image the skin surface of a patient in 3D before and during radiotherapy treatment. Part number: V000172 |
| Quantity | 690 systens |
| Distribution | Nationwide Foreign: Australia, Austria, Azerbaijan, Belgium, Brunei, Bulgaria, Canada, China, Croatia, Denmark, Finland, Georgia, Germany, India, Israel, Italy, Japan, Korea, Lithuania, Malaysia, Morocco, Norway, Poland, Portugal, Puerto Rico, Singapore, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Kingdom, Vietnam. |
| Initiated | 2018-08-24 |
| Classified | 2018-10-03 |
| Terminated | 2022-07-08 |
| Recall number | Z-0065-2019 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Vision RT Ltd (Z-0065-2019).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/81089/