RegWardenFDA facility compliance records
Recall

Class II — EOS Imaging

Recalling firmEOS imaging
Registered facilityEOS imaging (the firm's only registered facility)
ClassificationClass II
StatusTerminated
Reason3D projections may be incorrect when the pair of images used is a secondary capture generated from a series containing more than two images, where one of the images has been manually flipped. Projection errors may lead to errors in calculation of 3D clinical parameters when adjusting the position, size and shape of the 3D objects used. Incorrect clinical parameters can be a contributing factor in incorrect clinical decisions, or sub-optimal surgical interventions.
ProductsterEOS workstation
Quantity17 units
DistributionUS nationwide
Initiated2018-08-01
Classified2018-09-18
Terminated2022-08-15
Recall numberZ-3188-2018
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — EOS Imaging (Z-3188-2018).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/80938/

Not open yet

Alerts for EOS imaging

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.