RegWardenFDA facility compliance records
Recall

Class II — Johnson & Johnson Surgical Vision Inc

Recalling firmJohnson & Johnson Surgical Vision, Inc.
Registered facilityView this company's 9 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonSoftware upgrade exhibits failures when executing Daily Alignment Verification (DAV), specifically, cuts on the plastic hemisphere during DAV do not match the intended cuts shown on the treatment overlay screen on the system GUI.
ProductOptiMedica Catalys Precision Laser System, CE 0044 containing software cOS 5.00.33 Product Label for Serial Numbers: 44109015; 44106215; 44107315; and 44065613 label states "Manufactured for: OptiMedica Corporation 1310 Moffett Park Drive Sunnyvale, CA 94089 USA Product Label for Serial Numbers: 46120916; 46123116, 46124617; 46131418, and 46131618 label states "Manufactured for: OptiMedica Corporation 510 Cottonwood Drive Milpitas, CA 95035 USA The Catalys Precision Laser System is indicated for use in patients undergoing cataract surgery for removal of the crystalline lens. Intend…
Quantity9
DistributionUS Distribution to states of: CO, GA, IL, MD including Puerto Rico and internationally to countries of: Austria and Germany.
Initiated2018-05-09
Classified2018-10-03
Terminated2021-01-13
Recall numberZ-0066-2019
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Johnson & Johnson Surgical Vision Inc (Z-0066-2019).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/80849/

Not open yet

Alerts for Johnson & Johnson Surgical Vision, Inc.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.