RegWardenFDA facility compliance records
Recall

Class II — Intuitive Surgical, Inc.

Recalling firmIntuitive Surgical, Inc.
Registered facilityView this company's 41 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusOngoing
ReasonAn Endoscope Controller has been improperly calibrated during manufacturing. This improper calibration may result in the Endoscope Controller showing surgical images that are dimmer, have color imbalance or have reduced fluorescence and grayscale when using Firefly. There is also potential for increased fogging of the endoscope.
ProductThe da Vinci Xi Surgical System (IS4000) Model Name: IS4000, Endoscope Controller. Model Number: 372601-16.
Quantity1
DistributionU.S.
Initiated2018-05-17
Classified2018-08-16
Recall numberZ-2835-2018
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Intuitive Surgical, Inc. (Z-2835-2018).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/80691/

Not open yet

Alerts for Intuitive Surgical, Inc.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.