Recall

Class II — ConvaTec, Inc

Recalling firmConvaTec
Registered facilityView this company's 6 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonBubble Leak Testing on AbViser AutoValve IAP Monitoring Devices without Transducer confirmed the potential for a pinhole breach in the sterile barrier.
ProductAbViser AutoValve IAP (intra- abdominal pressure) Monitoring Device, Patient mount: a) REF ABV300, b) REF ABV301 AbViser is a sterile non-invasive disposable intra- abdominal pressure monitoring device containing aspiration tubing, infusion tubing, valves and optional pressure transducers for the measurement of intra- abdominal bladder pressure. The device attaches directly to the patient s existing urinary catheter/drain system providing both an enclosed fluid path for infusing fluid into the bladder catheter in order to monitor the hydrostatic pressure in the bladder.
Quantity3881 units
Distributionworldwide Distribution: US (nationwide) and countries of: Australia, Austria, Canada, Bahrain, Bangladesh, Belgium, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Estonia, Finland, France, Germany, Hungary, India, Israel, Italy, Kenya, Kuwait, Malta, Mauritius, Netherlands,Norway, Portugal, Qatar, Romania, Saudi Arabia, Singapore, Slovenia, South Africa, South Korea, Spain, Sweden, Swit…
Initiated2018-07-03
Classified2018-08-10
Terminated2023-01-24
Recall numberZ-2744-2018
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — ConvaTec, Inc (Z-2744-2018).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/80539/

Not open yet

Alerts for ConvaTec

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.