RegWardenFDA facility compliance records
Recall

Class II — CSL Behring GmbH

Recalling firmCSL Behring LLC
Registered facilityCSL Behring LLC (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonSiemens Healthcare Diagnostics has observed a negative bias for the carbohydrate-deficient transferrin (CDT) measurement when using the affected N Latex CDT lots in comparison to the HPLC method. CDT results in absolute concentrations may be influenced by patients transferrin levels and, therefore, results are reported as ratio of CDT to total transferrin, called %CDT. As the transferrin determination with N Antiserum to Human Transferrin is not affected, the observed effect leads to calculated %CDT values that show a negative bias of approximately 15% compared to the HPLC method. This could result in a shift of weak positive patient samples into the reference range of 1.19  2.47 %CDT which was derived from a study population of healthy adults. There are no control or calibrator materials that would catch this bias.
ProductN Latex CDT Kit
Quantity6635 units
DistributionDomestic - DE, IN, KS, MO & VA Internationally - Australia Austria Belgium Canada China Czech Republic Finland France Germany Hungary India Italy Japan Latvia Lithuania Netherlands Norway Poland Portugal Saudi Arabia Slovakia Slovenia South Africa South Koria Spain Sweden Switzerland Taiwan Turkey United Kingdom Uruguay
Initiated2018-05-31
Classified2018-08-25
Terminated2021-02-12
Recall numberZ-2931-2018
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — CSL Behring GmbH (Z-2931-2018).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/80431/

Not open yet

Alerts for CSL Behring LLC

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.