RegWardenFDA facility compliance records
Recall

Class II — Zeiss, Carl Inc

Recalling firmCarl Zeiss Inc.
Registered facilityView this company's 2 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonIn certain eyepiece configurations the laser shutter may not close completely, permitting reflected beams that may be greater than the Class I limit.
ProductCell Observer SD, DirectFRAP, Laser TIRF 1, Laser TIRF 2, Laser TIRF3, LightSheet, Elyra, and multiple Laser Scanning Microscopes. Laser scanning microscopes are used in cell biology research.
Quantity1070 units to US
DistributionUSA (nationwide)
Initiated2018-05-08
Classified2018-07-11
Terminated2020-04-21
Recall numberZ-2058-2018
InitiationFDA Mandated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Zeiss, Carl Inc (Z-2058-2018).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/80115/