RegWardenFDA facility compliance records
Recall

Class II — Think Surgical, Inc.

Recalling firmThink Surgical Inc.
Registered facilityThink Surgical Inc. (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonThere is a potential for the cutter head to detach from the cutter shaft.
ProductThink Surgical Cutter, Flat, 6.2mm x 144mm. For use only with the TCAT Computer-Assisted Tool. Part Number 106428 The cutter is used to prepare the bone for the (TKA) Total Knee Arthroplasty.
Quantity951
DistributionThe devices were distributed to the following US states: NJ, NY, OH, and TX. The devices were also distributed to Korea.
Initiated2018-03-30
Classified2018-08-03
Terminated2020-01-24
Recall numberZ-2605-2018
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Think Surgical, Inc. (Z-2605-2018).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/80091/