Class II — Draeger Medical, Inc.
| Recalling firm | Draeger Medical Systems, Inc. |
|---|---|
| Registered facility | View this company's 2 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | The valve assembly material may become corroded due to insufficient cleaning by a supplier. There is a risk of the breathing gas becoming contaminated in the affected devices. |
| Product | Carina Ventilator, ventilator. Catalog no. 5704110. Long-term ventilator for treatment in hospitals by a qualified medical personnel. |
| Quantity | 8 units |
| Distribution | Distributed to the states of CT, IL, MN, NV, SD, and WA. |
| Initiated | 2018-04-26 |
| Classified | 2018-05-29 |
| Terminated | 2018-09-07 |
| Recall number | Z-1954-2018 |
| Initiation | Voluntary: Firm initiated |