Class II — Genesee BioMedical, Inc.

Recalling firmGENESEE BIOMEDICAL, INC.
Registered facilityGENESEE BIOMEDICAL, INC. (the firm's only registered facility)
ClassificationClass II
StatusOngoing
ReasonFailure in a retractor blade due to corrosion. In the area of the blade where the bearing/button meets the blade body, corrosion can form, weakening the welded joint and may cause the bearing/button to break.
ProductSternal Valve Retractors With Atrial and Aortic Rakes; Models SRCL-1, SRCL-AR, SRLP-1, SRLP-AR, ATR-1, ATR-AR, CSR-1 and CSR-AR. For use in cardiac surgery.
Quantity444 units
DistributionWorldwide distribution. US nationwide, South Korea, Australia, COSTA RICA, Taiwan, New Zealand, Italy, Canada, Netherlands, Norway, Sweden, Switzerland, Israel, Japan, Egypt, South Africa, United Kingdom
Initiated2018-02-26
Classified2018-05-29
Recall numberZ-1953-2018
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Genesee BioMedical, Inc. (Z-1953-2018).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/80024/

Not open yet

Alerts for GENESEE BIOMEDICAL, INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.