Class II — Baxter Healthcare Corporation
| Recalling firm | BAXTER HEALTHCARE CORPORATION |
|---|---|
| Registered facility | View this company's 32 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | The firm will be updating the Instructions for Use for the product. Current IFUs instruct users to lower the primary bag 20 inches below the secondary bag using a hanger when performing a secondary infusion. However, the firm supplies 12.5-inch hangers with the Baxter disposable secondary sets. |
| Product | Baxter SIGMA Spectrum Infusion Pump with Master Drug Library (Version 8), infusion pump, 35700BAX2 Product Usage: The Baxter SIGMA Spectrum Infusion Pump with Master Drug Library is intended to be used for the controlled administration of fluids. These fluids may include pharmaceutical drugs, blood, blood products and mixture of required patient therapy. The intended routes of administration consist of the following clinically accepted routes: intravenous, arterial, subcutaneous, intrathecal, epidural or irrigation of fluid space. Spectrum Infusion Pump is intended to be used in conjunctio… |
| Quantity | 122889 |
| Distribution | Worldwide Distribution - US Nationwide and the country of Canada. |
| Initiated | 2018-05-08 |
| Classified | 2018-06-07 |
| Terminated | 2019-04-08 |
| Recall number | Z-2113-2018 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Baxter Healthcare Corporation (Z-2113-2018).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/80014/
Alerts for BAXTER HEALTHCARE CORPORATION
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.