RegWardenFDA facility compliance records
Recall

Class II — Orthofix, Inc

Recalling firmORTHOFIX SRL
Registered facilityORTHOFIX SRL (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonFirm is conducting a voluntary recall of the Rival Reduce Fracture Plating System and Rival View Plating System. As the result of a complaint investigation, the locking screws were confirmed to not properly engage the plates during surgery.
ProductLapidus Plate, Sterile, Rx Only, ORTHOFIX S.r.r. Italy on label, Size/Step/Left or Right (Part Number): a) Small/0MM Step/Left (002-C-01000L); b) Small/0MM Step/Right (002-C-01000R); c) Large/0MM Step/Left (002-C-02000L); d) Large/0MM Step/Right (002-C-02000R); e) Small/2MM Step/Left (002-C-01002L); f) Small/2MM Step/Right (002-C-01002R); g) Large/2MM Step/Left (002-C-02002L); h) Large/2MM Step/Right (002-C-02002R)
Quantity156
DistributionU.S Consignees: NJ, FL, PA, SC, LA, VA, MN, TX, GA, DC, NV, IN, CA, MI, IL, MD, NC, OH, OK,
Initiated2018-04-06
Classified2018-06-07
Terminated2021-06-30
Recall numberZ-2129-2018
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Orthofix, Inc (Z-2129-2018).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/79819/