RegWardenFDA facility compliance records
Recall

Class II — Siemens Medical Solutions USA, Inc

Recalling firmSIEMENS MEDICAL SOLUTIONS USA, INC.
Registered facilityView this company's 9 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonCertain systems may have leaking of the plastic coolant couplings, the liquid can enter into the flat panel detector leading to a failure of the detector. It may be necessary to cancel or restart the treatment or transfer the patient to an alternate or another system.
ProductAXIOM Artis angiographic x-ray system, Model Numbers: 7412807, 7413078, 7728392 AXIOM Artis is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.
Quantity22
DistributionUS Distribution in the following U.S. states: AZ, CA, FL, IA, IL, LA, MI, NC, NJ, OH, OK, and TX.
Initiated2018-01-05
Classified2018-03-21
Terminated2018-08-23
Recall numberZ-1150-2018
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Siemens Medical Solutions USA, Inc (Z-1150-2018).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/79540/