Class II — Smith & Nephew, Inc.
| Recalling firm | Smith & Nephew, Inc. |
|---|---|
| Registered facility | View this company's 15 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | One lot of D-RAD Self-Taping Locking Screws used with the Distal Radius Fracture Kit were machined out of specification. Screws measured over tolerance within the head thread form by 0.002-0.003 inch. |
| Product | D-RAD Self-Tapping Locking Screw intended for the fixation of fractures involving the distal radius. smith&nephew D-RAD 2.4MM X 10MM S-T LOCKING SCREW, REF 74692410, QTY (1), STERILE R |
| Quantity | 59 units |
| Distribution | US, Puerto Rico, Malaysia |
| Initiated | 2018-03-07 |
| Classified | 2018-04-06 |
| Terminated | 2018-10-31 |
| Recall number | Z-1317-2018 |
| Initiation | Voluntary: Firm initiated |