RegWardenFDA facility compliance records
Recall

Class II — Smith & Nephew, Inc.

Recalling firmSmith & Nephew, Inc.
Registered facilityView this company's 15 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonOne lot of D-RAD Self-Taping Locking Screws used with the Distal Radius Fracture Kit were machined out of specification. Screws measured over tolerance within the head thread form by 0.002-0.003 inch.
ProductD-RAD Self-Tapping Locking Screw intended for the fixation of fractures involving the distal radius. smith&nephew D-RAD 2.4MM X 10MM S-T LOCKING SCREW, REF 74692410, QTY (1), STERILE R
Quantity59 units
DistributionUS, Puerto Rico, Malaysia
Initiated2018-03-07
Classified2018-04-06
Terminated2018-10-31
Recall numberZ-1317-2018
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Smith & Nephew, Inc. (Z-1317-2018).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/79515/