Recall

Class II — Siemens Medical Solutions USA, Inc

Recalling firmSIEMENS MEDICAL SOLUTIONS USA, INC.
Registered facilityView this company's 9 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
Reasonto inform customers of possible incorrect tube current calculations by the CARE Dose4D algorithm for head scans based on p.a. (posterior-anterior) or a.p. (anterior-posterior) topograms. Depending on the geometrical shape of the skull bone, it may happen in rare cases that the calculated dose distribution is not appropriate and could lead to unnecessary radiation exposure
ProductSOMATOM Perspective 16 Intended to produce cross-sections images of the body by computer reconstruction of x-ray transmission data.
Quantity1301
DistributionUSA (nationally) Distribution.
Initiated2018-02-15
Classified2018-04-30
Terminated2020-04-08
Recall numberZ-1448-2018
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Siemens Medical Solutions USA, Inc (Z-1448-2018).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/79428/

Not open yet

Alerts for SIEMENS MEDICAL SOLUTIONS USA, INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.