Class II — Medica Corporation

Recalling firmMEDICA CORP.
Registered facilityMEDICA CORP. (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonThe firm has initiated a field recall of the Wash1 Kit REF 10680 due to reported inconsistent Lipase values when analyzing patient samples. The firm's internal investigation identified the underlying cause to be an ineffective wash solution resulting in an interference carryover from the triglyceride (TRIG) assay if they are on the same worklist.
ProductWash Kit (REF 10680) Lots 16279, 17045 and 17215 Medica corporation is marketing a wash solution Chemistry Wash1 , REF # 10680 for use with Medica s EasyRA clinical chemistry analyzer. The solution is used prior to a lipase (LIP) assay if an interferant assay (TRIG)is part of the work list. The major interferant for Lipase is the Triglycerides (TRIG) assay since one of the reagents contains lipase as an active ingredient.
Quantity392
DistributionWorldwide and US Distribution.
Initiated2017-11-21
Classified2018-03-21
Terminated2020-08-20
Recall numberZ-1149-2018
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Medica Corporation (Z-1149-2018).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/79264/

Not open yet

Alerts for MEDICA CORP.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.