RegWardenFDA facility compliance records
Recall

Class II — Fresenius Medical Care Renal Therapies Group, LLC

Recalling firmFresenius Medical Care Renal Therapies Group, LLC
Registered facilityView this company's 11 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonWhile reviewing documentation for the next software release of the 2008T, an R&D technician identified that setting the UF goal to '0' introduces a discrepancy between the UF rate displayed and the actual UF pump rate. The software anomaly is also applicable to the 2008T Machines that contained the 2.63 Bug Fix per DCAF 17-088. 2008T Upgrade kits that contained the 2.63 Bug Fix are also affected. Additionally, SW version 2.64 (in design freeze) is also impacted by this anomaly.
Product2008T, Hemodialysis Delivery System Product Indicated for acute and chronic dialysis therapy.
Quantity15 machines
DistributionUS Nationwide Distribution
Initiated2017-10-05
Classified2018-03-01
Terminated2021-07-28
Recall numberZ-0829-2018
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Fresenius Medical Care Renal Therapies Group, LLC (Z-0829-2018).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/79239/

Not open yet

Alerts for Fresenius Medical Care Renal Therapies Group, LLC

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.