Class II — Ion Beam Applications S.A.
| Recalling firm | ION BEAM APPLICATIONS S.A. |
|---|---|
| Registered facility | ION BEAM APPLICATIONS S.A. (the firm's only registered facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | Range for the same energy could be different depending if the layer is the first layer of the irradiation plan or not. |
| Product | Proteus 235, Pencil Beam Scanning version PTS-6.4.11.X The Proton Therapy System - Proteus 235 (brand names: Proteus Plus and Proteus ONE) is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation. The PTS may include a fixed small beam treatment room dedicated to the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation localized to the head and neck. |
| Quantity | 3 |
| Distribution | US Distribution to the state of :FL |
| Initiated | 2017-10-19 |
| Classified | 2018-03-08 |
| Terminated | 2019-04-02 |
| Recall number | Z-0931-2018 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Ion Beam Applications S.A. (Z-0931-2018).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/79228/
Alerts for ION BEAM APPLICATIONS S.A.
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.