Class II — Covidien LLC
| Recalling firm | Covidien llc |
|---|---|
| Registered facility | View this company's 22 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | Potential for failure of the device safety interlock. The safety interlock prevents an empty single-use loading unit from being fired a second time. |
| Product | Endo GIA Auto Suture Universal Articulating Loading Unit for use in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection and creation of anastomosis. |
| Quantity | 3,816 |
| Distribution | Nationally |
| Initiated | 2017-09-13 |
| Classified | 2018-02-16 |
| Terminated | 2019-06-28 |
| Recall number | Z-0636-2018 |
| Initiation | Voluntary: Firm initiated |