RegWardenFDA facility compliance records
Recall

Class II — BioMerieux SA

Recalling firmBIOMERIEUX SA
Registered facilityView this company's 16 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonElute may become colored due to the residual presence of heme group origination from hemoglobin present in whole blood samples as well as dried spot (DBS) samples. This can potentially cause the inhibition of PCR reactions and result in an uninterpretable test result. If using an Internal Control (IC) in the extraction process the IC could potentially be inhibited as well and invalidate test results.
ProductNucliSENS Lysis Buffer Extended lot for RES 76675
Quantity911 units
DistributionDistributed domestically to MD, CO, and OH.
Initiated2017-10-18
Classified2017-11-09
Terminated2021-08-24
Recall numberZ-0086-2018
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — BioMerieux SA (Z-0086-2018).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/78387/