RegWardenFDA facility compliance records
Recall

Class II — Firefly Medical, Inc.

Recalling firmFIREFLY MEDICAL, INC.
Registered facilityFIREFLY MEDICAL, INC. (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonThe dowel pins (P/N 16082) that fit in the left and right leg pivot castings (P/Ns 16049/16056) can rotate and gradually maneuver out of the leg casting after several hundred deployment and folding cycles (designed for easy storage) of the IVEA unit.
ProductIVEA Product Part Number 16100 Model 500A
Quantity570 units
DistributionUnited States, Australia, Belgium, Canada, China, Germany, Hong Kong, Indonesia, Israel, Netherlands, Singapore, UAE
Initiated2017-08-24
Classified2017-09-29
Terminated2019-11-01
Recall numberZ-3239-2017
InitiationVoluntary: Firm initiated