Class II — Biomerieux Inc
| Recalling firm | BIOMERIEUX SA |
|---|---|
| Registered facility | View this company's 16 registered facilities → (openFDA recall records do not name a specific facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | Customers have reported that some VITEK® 2 cards are staying in preliminary status, not finalizing after ejection from the instrument, and not allowing cards in subsequent carousel slots to be processed. The issue was reported to occur on VITEK® 2 Compact 15 and Compact 30 systems following a system software update to version 8.01. |
| Product | VITEK® 2 Compact 30, REF numbers: 27530 and 27530R (clinical), and 27630 and 17630R (industry) software and reagent cards designed for the identification (ID) and antimicrobial susceptibility testing (AST) of bacteria and yeast. |
| Quantity | 9234 units |
| Distribution | Worldwide |
| Initiated | 2017-07-12 |
| Classified | 2017-08-25 |
| Terminated | 2020-12-16 |
| Recall number | Z-3076-2017 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record