Class II — Solta Medical Inc

Recalling firmSolta Medical
Registered facilitySolta Medical, Inc. (the firm's only registered facility)
ClassificationClass II
StatusTerminated
ReasonPotential inaccurate calibration by service depot repair, which could contribute to patients being burnt by the device. This only impacts units serviced between the dates of 09/08/2016 and 05/23/2017.
ProductVASER PRO Amplifier, VASER 2.2 VASERlipo System stem, Rx Only, 115/230V - 60/50Hz 300VA The Vaser amplifier is intended for the fragmentation and emulsification of subcutaneous fatty tissues for aesthetic body contouring (modification of the anatomy).
Quantity7 units
DistributionUS Distribution to the states of : AL, CA, CO, FL, NY, PA, & TX.
Initiated2017-06-01
Classified2017-08-21
Terminated2020-11-30
Recall numberZ-3048-2017
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Solta Medical Inc (Z-3048-2017).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/77655/

Not open yet

Alerts for Solta Medical

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.