Class II — Implant Direct Sybron Manufacturing, LLC
| Recalling firm | Implant Direct |
|---|---|
| Registered facility | IMPLANT DIRECT SYBRON MANUFACTURING LLC (the firm's only registered facility) |
| Classification | Class II |
| Status | Terminated |
| Reason | The inside diameter of the 2.3mm small drill guide inserts have a smaller diameter than the specifications require. |
| Product | Spectra-System/Legacy Guided Surgery Handle Kit, Device Listing Number D245883. The Guided Handle Kit (G-HK) allows clinicians to drill through a surgical guide to create an osteotomy for implant placement with the precision and accuracy as designed with treatment planning software. The G-HK consist of 4 guide handles with inserts attached on each end. Each insert has an opening that matches the diameter of our surgical drills. The insert portion of the guide handle fits intimately into the guide sleeve that is embedded into a surgical guide. |
| Quantity | 84 units |
| Distribution | Worldwide Distribution - USA (nationwide) Distribution to the states of : AZ, CA, CO, CT, FL, ID, IL, KS, MI, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OR, PA, SC, TN, TX, UT, VA, WA., and to the countries of : Canada, Netherlands, Spain, Great Britain, Germany and Switzerland. |
| Initiated | 2017-06-26 |
| Classified | 2017-09-11 |
| Terminated | 2021-06-16 |
| Recall number | Z-3115-2017 |
| Initiation | Voluntary: Firm initiated |
Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)
Cite this record
RegWarden. “Class II recall — Implant Direct Sybron Manufacturing, LLC (Z-3115-2017).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/77554/
Alerts for Implant Direct
Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.
Want a one-page facility brief?
A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.