RegWardenFDA facility compliance records
Recall

Class II — Becton Dickinson & Co.

Recalling firmBecton Dickinson and Company
Registered facilityView this company's 36 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
ReasonBD has recently confirmed quality control failures, related to a manufacturing issue, for the inducible macrolide-linosamide-streptogramin B (iMLSb) test contained on lot 7031818 of Phoenix panels. The iMLSb resistance D-test screen for Staphylococcus species is used to identify phenotypic drug resistance to clindamycin and erythromycin. As the product should not be used when quality control (QC) fails, it is unlikely that there would be any impact on patient results, however, if a customer does not perform QC testing and uses the panel, there is a potential to incorrectly treat a Staphylococcus infection with clindamycin based on the iMLSb test result. This issue can be detected 100% of the time when used per package insert instructions with iMLSb positive strain S. aureus BAA-977.
ProductBD Phoenix PMIC-107, Catalog Number 448417, UDI #s (01)30382904484171, (10)7031818(17)180228(30)0025, IVD. PMIC-107 is a Gram positive AST Only panel used to test antibiotic susceptibility. The inducible macrolide-linosamide-steptogramin B (iMLSb) resistance D-test screen for Staphylococcus species is used to identify phenotypic drug resistance to clindamycin and erythromycin. --- Product is packaged 25 panels to 1 box.
QuantityUS: 7750 units
DistributionCT, VA, PA, TX, FL, NC, WI, NM
Initiated2017-05-11
Classified2017-11-25
Terminated2018-02-06
Recall numberZ-0171-2018
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Becton Dickinson & Co. (Z-0171-2018).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/77370/

Not open yet

Alerts for Becton Dickinson and Company

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.