Class II — Zimmer Gmbh

Recalling firmZIMMER, INC.
Registered facilityView this company's 10 registered facilities → (openFDA recall records do not name a specific facility)
ClassificationClass II
StatusTerminated
Reasonthe devices used a different raw material to produce two countersink products which has higher hardness and higher carbon content resulting in device damage.
ProductNormed General Instruments 503004541: Werber Countersink Cannulated for micro CBS Screws, AO 503004341: Werber Countersink Cannulated for micro CBS Screws, round-shaft The Countersink is an instrument which is used over the guide wire to prepare for adequate space in the cortical bone rim to sink the screw head in to bones/tissues for different Foot, Ankle and Hand implant systems. The instrument is used in several Normed implant systems. The instrument is used to avoid soft tissue irritation
Quantity138 (OUS 297)
DistributionSingle US consignee CO.
Initiated2017-05-11
Classified2017-05-18
Terminated2017-11-03
Recall numberZ-2111-2017
InitiationVoluntary: Firm initiated

Source: U.S. FDA via openFDA (opens in a new tab) (device registration & enforcement). Data as of 2026-10-06. Verify this recall record at FDA → (opens in a new tab)

Cite this record

RegWarden. “Class II recall — Zimmer Gmbh (Z-2111-2017).” Compiled from U.S. FDA public records (openFDA; FDA Data Dashboard). Data as of 2026-10-06. https://regwarden.com/recall/77262/

Not open yet

Alerts for ZIMMER, INC.

Watchlist alerts will email you when FDA publishes a new recall, inspection classification, or warning letter for a facility you follow. They aren't open to the public yet — leave your email and we'll tell you when they are.

Planned

Want a one-page facility brief?

A single-page summary of this facility's registration and recall history for diligence. Not selling it yet — register interest and we'll follow up.